Dosimetric evaluation in Helical TomoTherapy for lung SBRT using Monte Carlo-based independent dose verification software

CONCLUSION: The TPS of HT might overestimate the PTV dose by more than 5% if any the three indices in this study were below threshold. Therefore, independent verification with an MC-based algorithm should be strongly recommended for lung SBRT in HT.PMID:38368607 | DOI:10.1002/acm2.14305
Source: Journal of Applied Clinical Medical Physics - Category: Physics Authors: Source Type: research