Impact of log file source and data frequency on accuracy of log file-based patient specific quality assurance

This study compared the dosimetric accuracy of high-frequency linear accelerator (Linac) log files and low-frequency log data stored in the oncology information system (OIS). Thirty patients were included, that were recently treated in the head and neck (HN), brain, and prostate region with volumetric modulated arc therapy (VMAT) and an additional ten patients treated using stereotactic body radiation therapy (SBRT) with 3D-conformal radiotherapy (3D-CRT) technique. Log data containing a single fraction were used to calculate the dose distributions. The dosimetric differences between Linac log files and OIS logs were evaluated with a gamma analysis with 2%/2 mm criterion and dose threshold of 30%. The original treatment plan was used as a reference. Moreover, DVH parameters of D98%, D50%, and D2% of the planning-target volume (PTV) and dose to several organs at risk (OARs) were reported. Significant differences in dose distributions between the two log types and the original dose were observed for PTV D98% and D2% (r < 0.001) for HN cases, PTV D98% (r = 0.005) for brain cases, and PTV D50% (r = 0.015) for prostate cases. No significant differences were found between the two log types with respect to D50%. The root mean square (RMS) error of the leaf positions of the OIS log was approximately twice the RMS error of the Linac log file for VMAT plans, but identical for 3D-CRT plans. The relationship between the gamma pass rate and the RMS error showed a moderate correlation f...