Monte Carlo dose calculation & #8211; A QA method for srt and sbrt plans in treating multiple and small metastatic lesions

Teh Lin, Lu Wang, C-M Charlie MaJournal of Medical Physics 2022 47(1):99-104To provide accurate and fast 3-D dose verification for hypofractionated stereotactic radiotherapy (SRT/SBRT) of small and multi targets calculated with a Varian Eclipse treatment planning system (TPS) delivered on a Varian accelerator. Ten brain and lung hypofractionated SRT/SBRT linac-based and CyberKnife plans were generated by the Eclipse system for delivery on the accelerator with the Millenium-120 leaf multileaf collimator (MLC) and Multiplan for the CyberKnife machine. These clinical SRT/SBRT plans required accurate quality assurance measurements to obtain absolute point dose and 3-D dose distributions due to the low number of fractions and high fraction doses. For small-field and multi-target plans, the EGS4/MCSIM code was used to calculate the dose distribution. A 0.125 cc ion chamber, a 0.016 cc pin-point chamber and Kodak EDR2 film were used for the measurements and the results were compared with Monte Carlo (MC) calculations. The dosimetry for small-field and multi-target treatment plans is challenging due to the comparable range of secondary electrons and the field sizes defined by SRT/SBRT MLC segments. Our MC simulations can accurately reproduce the linac dose distributions (within 1%/1 mm) three dimensionally. For the clinical SRT/SBRT plans investigated in this work, the MC doses agreed within 3% with ion chamber measurements and within 2%/2 mm with fil...
Source: Journal of Medical Physics - Category: Physics Authors: Source Type: research
More News: Brain | Neurology | Physics