A novel angular dependency model for MatriXX response and its application to true composite dose verification for IMRT plans
CONCLUSION: The proposed model for angular dependent MatriXX response is necessary and effective. Dose verification using MatriXX with angular-model-based approach is acceptable for true composite beam IMRT plans with required accuracy to simplify patient specific QA.PMID:34453876 | DOI:10.1002/acm2.13405