Steadily Increasing Inversion Time Improves Blood Suppression for Free-Breathing 3D Late Gadolinium Enhancement MRI With Optimized Dark-Blood Contrast

Objectives Materials and Methods Fifty consecutive patients with previous cardiac arrhythmias, scheduled for high-resolution 3D LGE MRI, were prospectively enrolled between October 2017 and February 2020. Free-breathing 3D dark-blood LGE MRI with high isotropic resolution (1.6 × 1.6 × 1.6 mm) was performed using a conventional fixed TI (n = 25) or a dynamic TI (n = 25). The average increase in blood nulling TI per minute was obtained from Look-Locker scans before and after the 3D acquisition in the first fixed TI group. This average increment in TI was used as input to calculate the dynamic increment of the initial blood nulling TI value as set in the second dynamic TI group. Regions of interest were drawn in the left ventricular blood pool to assess mean signal intensity as a measure for blood pool suppression. Overall image quality, observer confidence, and scar demarcation were scored on a 3-point scale. Results Three-dimensional dark-blood LGE data sets were successfully acquired in 46/50 patients (92%). The calculated average TI increase of 2.3 ± 0.5 ms/min obtained in the first fixed TI group was incorporated in the second dynamic TI group and led to a significant decrease of 72% in the mean blood pool signal intensity compared with the fixed TI group (P
Source: Investigative Radiology - Category: Radiology Tags: Technical Note Source Type: research