Real-Time Gain Control of PET Detectors and Evaluation With Challenging Radionuclides

We describe the methods used to combine information about multiple peaks and how this algorithm is implemented in a way that permits real-time processing on a field-programmable gate array. Simulations demonstrate rapid response time and stability. A method (“virtual scatter filter”) is also described that extracts unscattered photopeak events from phantom data and demonstrates the accuracy of the photopeak for various radionuclides that emit energies in addition to the pure 511 keV annihilation peak. Radionuclides 52 Mn, 55 Co, 64 Cu, 89 Zr, 90 Y, and 124 I are included in the study for their various forms of spectral contamination.
Source: IEE Transactions on Medical Imaging - Category: Biomedical Engineering Source Type: research