Recognizing and Correcting MEG Artifacts

Summary: Noise sources in magnetoencephalography (MEG) include: (1) interference from outside the shielded room, (2) other people and devices inside the shielded room, (3) physiologic or nonphysiologic sources inside the patient, (4) activity from inside the head that is unrelated to the signal of interest, (5) intrinsic sensor and recording electronics noise, and (6) artifacts from other apparatus used during recording such as evoked response stimulators. There are other factors which corrupt MEG recording and interpretation and should also be considered “artifacts”: (7) inadequate positioning of the patient, (8) changes in the head position during the recording, (9) incorrect co-registration, (10) spurious signals introduced during postprocessing, and (11) errors in fitting. The major means whereby magnetic interference can be reduced or eliminated are by recording inside a magnetically shielded room, using gradiometers that measure differential magnetic fields, real-time active compensation using reference sensors, and postprocessing with advanced spatio-temporal filters. Many of the artifacts that plague MEG are also seen in EEG, so an experienced electroencephalographer will have the advantage of being able to transfer his knowledge about artifacts to MEG. However, many of the procedures and software used during acquisition and analysis may themselves contribute artifact or distortion that must be recognized or prevented. In summary, MEG artifacts are not worse t...
Source: Journal of Clinical Neurophysiology - Category: Neurology Tags: Invited Review Source Type: research