Operative and Technical Modifications to the Coriolis ® µ Air Sampler That Improve Sample Recovery and Biosafety During Microbiological Air Sampling

We report inconsistent sample recovery from the collector cone due to loss of material when sampling continuously for more than 30 min. Introducing a new collector cone every 10 min improved this shortcoming. Moreover, we found that several surfaces on the device become contaminated during sampling. Adapting a high efficiency particulate air-filter system to the Coriolis prevented contamination without altering collection efficiency or tactical deployment. A Coriolis modified with these operative and technical improvements was used to collect aerosols carrying microspheres released inside a Biosafety Level-3 laboratory during simulations of microbiological spills and aerosol dispersals. In summary, we provide operative and technical solutions to the Coriolis that optimize microbiological air sampling and improve biosafety.
Source: Annals of Occupational Hygiene - Category: Occupational Health Source Type: research