Comparison of Three Blood Collection Tubes for 35 Biochemical Analytes: The Becton Dickinson Barricor Tube, Serum Separating Tube, and Plasma Separating Tube.

Comparison of Three Blood Collection Tubes for 35 Biochemical Analytes: The Becton Dickinson Barricor Tube, Serum Separating Tube, and Plasma Separating Tube. Ann Lab Med. 2021 01;41(1):114-119 Authors: Shin S, Oh J, Park HD Abstract The Barricor tube (Becton Dickinson [BD], Sunnyvale, CA, USA) was recently developed to mechanically separate plasma by increasing the centrifugation rate. We compared the Barricor tube with existing serum- and plasma-based tubes based on 35 biochemical analytes and preanalytical turnaround time (TAT). Blood samples were collected from 30 healthy volunteers in a Barricor tube, serum separating tube (SST, Vacutainer SST II Tube 8.5 mL, #368972; BD), or plasma separating tube (PST, Vacutainer PST Tube 8.0 mL, #367964; BD) in random order. Next, 27 chemistry analytes, six immunochemistry analytes, and two cardiac markers were compared using Passing-Bablok regression and the Bland-Altman method. Preanalytical TAT was measured for each tube. The Barricor tube exhibited bias exceeding the desirable limit for nine and four analytes compared with the SST and PST, respectively. The Barricor tube lactate dehydrogenase value showed a bias of -10.29% and -9.86% compared with that of the SST and PST, respectively. The preanalytical TAT of Barricor tube was 8.8 minutes, which was the shortest among the three tubes. The clinical performance of the Barricor tube was equivalent to that of the SST and PST for most analyte...
Source: Annals of Laboratory Medicine - Category: Laboratory Medicine Tags: Ann Lab Med Source Type: research