Characterization of RDX and HMX explosive adduct ions using ESI FT ‐ICR MS

AbstractInvestigation of two common explosives such as cyclonite (RDX) and cyclotetramethylenetetranitramine (HMX) using a mass spectrometer with ultrahigh resolution and accuracy hasn't been comprehensively performed. Here, ultrahigh mass accuracy 15 ‐T Fourier transform‐ion cyclotron resonance mass spectrometry (FT‐ICR MS) spectra were utilized to comprehensively characterize the adduct ions of RDX and HMX. Two different ionization sources such as a conventional electrospray ionization (ESI) source and a chip‐based static nano‐ESI sou rce were used to investigate the adduct ions of RDX and HMX. The ESI‐MS analyses of two explosives in negative ion mode provides some adduct ions of RDX and HMX even without prior addition of their corresponding anions. A total of six types of adduct ion were characterized: [M + Cl]‐, [M + HCOO]‐, [M + NO2]‐, [M + CH3COO]‐, [M + NO3]‐, and [M + C3H5O3]‐, where M is either RDX or HMX. The ultrahigh accuracy of the 15 ‐T FT‐ICR MS was utilized to distinguish two closely spaced peaks representing the monoisotopic [M + NO2]‐ and second isotopic [M + HCOO]‐ ions, thereby enabling the discovery of a [M + NO2]‐ adduct ion in the ESI analysis of RDX or HMX. [M + NO2]‐ and [M + CH3COO]‐ adduct ions were only observed when using a static nano ‐ESI source. It is the first report explaining the discovery of [M + NO2]– adduct ion in the ESI ‐MS analyses of RDX and HMX.
Source: Journal of Mass Spectrometry - Category: Chemistry Authors: Tags: SPECIAL ISSUE ‐ RESEARCH ARTICLE Source Type: research