Towards improving the accuracy of aortic transvalvular pressure gradients: rethinking Bernoulli

AbstractThe transvalvular pressure gradient (TPG) is commonly estimated using the Bernoulli equation. However, the method is known to be inaccurate. Therefore, an adjusted Bernoulli model for accurate TPG assessment was developed and evaluated. Numerical simulations were used to calculate TPGCFD in patient-specific geometries of aortic stenosis as ground truth. Geometries, aortic valve areas (AVA), and flow rates were derived from computed tomography scans. Simulations were divided in a training data set (135 cases) and a test data set (36 cases). The training data was used to fit an adjusted Bernoulli model as a function of AVA and flow rate. The model-predicted TPGModel was evaluated using the test data set and also compared against the common Bernoulli equation (TPGB). TPGB and TPGModel both correlated well with TPGCFD (r >  0.94), but significantly overestimated it. The average difference between TPGModel and TPGCFD was much lower: 3.3  mmHg vs. 17.3 mmHg between TPGB and TPGCFD. Also, the standard error of estimate was lower for the adjusted model: SEEModel = 5.3 mmHg vs. SEEB = 22.3 mmHg. The adjusted model’s performance was more accurate than that of the conventional Bernoulli equation. The model might help to improve non-invasive assessment of TPG.Graphical abstractProcessing pipeline for the definition of an adjusted Bernoulli model for the assessment of transvalvular pressure gradient. Using CT image data, the patient specific geometry of the st...
Source: Medical and Biological Engineering and Computing - Category: Biomedical Engineering Source Type: research