Assessing the fit of 3D printed bolus from CT, optical scanner and photogrammetry methods

This study quantitatively compares the fits of boluses designed from different imaging modalities. A head phantom was imaged using three systems: a CT simulator, a 3D optical scanner, and an interchangeable lens camera. Nose boluses were designed and 3D printed from each modality. A 3D printed phantom with air gaps of known thicknesses was used to calibrate mean HU to measure air gaps of unknown thickness and assess the fit of each bolus on the head phantom. The bolus created from the optical scanner data resulted in the best fit, with a mean air gap of 0.16  mm. Smoothing of the CT bolus resulted in a more clinically suitable model, comparable to that from the optical scanner method. The bolus produced from the photogrammetry method resulted in air gaps larger than 1 mm in thickness. The use of optical scanner and photogrammetry models have many adva ntages over the conventional bolus-from-CT method, however workflow should be refined to ensure accuracy if implemented clinically.
Source: Australasian Physical and Engineering Sciences in Medicine - Category: Biomedical Engineering Source Type: research