Dosimetric verification of IMPT using a commercial heterogeneous phantom.

Dosimetric verification of IMPT using a commercial heterogeneous phantom. J Appl Clin Med Phys. 2019 Jan 23;: Authors: Yasui K, Toshito T, Omachi C, Hayashi K, Kinou H, Katsurada M, Hayashi N, Ogino H Abstract The purpose of this study was to propose a verification method and results of intensity-modulated proton therapy (IMPT), using a commercially available heterogeneous phantom. We used a simple simulated head and neck and prostate phantom. An ionization chamber and radiochromic film were used for measurements of absolute dose and relative dose distribution. The measured doses were compared with calculated doses using a treatment planning system. We defined the uncertainty of the measurement point of the ionization chamber due to the effective point of the chamber and mechanical setup error as 2 mm and estimated the dose variation base on a 2 mm error. We prepared a HU-relative stopping power conversion table and fluence correction factor that were specific to the heterogeneous phantom. The fluence correction factor was determined as a function of depth and was obtained from the ratio of the doses in water and in the phantom at the same effective depths. In the simulated prostate plan, composite doses of measurements and calculations agreed within ±1.3% and the maximum local dose differences of each field were 10.0%. Composite doses in the simulated head and neck plan agreed within 4.0% and the maximum local dose difference for...