Evaluation of Diagnostic Tests by Evanescence Biosensor Technology for Rapid Phenotyping of the Human Platelet Alloantigens 1a and 5b

Background: The human platelet alloantigens (HPA) HPA-1a and HPA-5b are located on glycoproteins on the platelet surface and are the most relevant to cause neonatal alloimmune thrombocytopenia (NAIT). The antigens are defined by single nucleotide polymorphisms (SNPs) in the glycoprotein genes, and the antigen status can be determined by genotyping the SNPs. However, genotyping is time-consuming and costly depending on the method and sample throughput. Here, we tested the reliability of the evanescence wave based fluorescence (EVA) biosensor technology for the rapid phenotyping of the HPA-1a and HPA-5b antigens on blood donor samples in two laboratories.Methods: HPA-1a and HPA-5b phenotyping was performed on EDTA blood samples from 336 blood donors (Lyon: 216 donors; Mannheim: 120 donors) using EVA typing assays and the biosensor system (Davos Diagnostics, Davos, Switzerland). For genotyping, validated PCR-SSP and TaqMan-PCR methods were used.Results: HPA-1a phenotyping was positive for all samples with HPA-1aa (n = 244; EVA value 807 ± 167 U/s) and HPA-1ab (n = 82; 542 ± 110 U/s) genotypes. All samples (n = 10) with negative EVA values (