A comparison of two pencil beam scanning treatment planning systems for proton therapy.
CONCLUSION: Both Eclipse and Raystation PB TPS are not always accurate within ±3% for a 5 cm range shifters or for small targets. This was improved with the Raystation MC model. The point dose calculations of Eclipse, Raystation PB, and Raystation MC compare within ±3% to measured doses for the other scenarios tested.
PMID: 29205763 [PubMed - as supplied by publisher]
Source: Journal of Applied Clinical Medical Physics - Category: Physics Authors: Langner UW, Mundis M, Strauss D, Zhu M, Mossahebi S Tags: J Appl Clin Med Phys Source Type: research