Quality assurance and average glandular dose measurement in mammography units

This study evaluated different operating parameters through mechanical test, accelerating voltage (kVp) accuracy test, machine output measurement, half-value layer measurement, calibration of compression device, image quality assessment, measurement of leakage radiation, radiation survey, and average glandular dose (AGD) measurements using stereotactic needle biopsy phantom. The results show that out of eight mammography units, only a single mammography unit (U-1) passed all QA tests and 2 units passed 7 tests, 2 units passed 6 tests, and 3 units passed 5 tests out of 8 QA tests. In unit 5, the AGD value was 4 and 1.93 mGy before and after service, respectively. QA programs as recommended by AAPM and AERB should be carried out periodically to ensure safety in breast cancer screening. This work points to the importance of the regulation and effective compliance and also help in both improving the QA and reduce the glandular dose received by the patients.