Early Hemodynamic Results in Patients With Small Aortic Annulus After Aortic Valve Replacement

Objective: Patients with a small aortic annulus (≤21 mm) have an increased risk of patient-prosthesis mismatch after valve replacement. The aim of this study was to compare the early hemodynamic performance of the balloon-expandable transaortic valve implantation Edwards system (SAPIEN) and the sutureless Perceval prostheses. Methods: Fifty patients underwent transcatheter aortic valve implantation, and 113 patients underwent sutureless aortic valve replacement. Mean ± SD aortic annulus diameter was 19.7 ± 1 mm, with no significant difference between groups. SAPIEN valve size was 23 mm in 40 patients (80%) and 26 mm in 10 patients (20%). Perceval valve size was small in 45 patients (40%), medium in 62 patients (55%), and large in 6 patients (5%). Transthoracic Doppler echocardiographic images were collected at baseline and before discharge. Results: There were no significant difference in predischarge effective orifice area (SAPIEN: 1.5 ± 0.5 cm2 and Perceval: 1.48 ± 0.34 cm2, P = 0.58) and indexed effective orifice areas (SAPIEN: 0.93 ± 0.32 cm2/m2 and Perceval: 0.88 ± 0.22 cm2/m2, P = 0.42). Predischarge mean ± SD transaortic gradient was lower with the SAPIEN than with Perceval valves (12 ± 6 and 17 ± 6 mm Hg, respectively, P
Source: Innovations: Technology and Techniques in Cardiothoracic and Vascular Surgery - Category: Cardiovascular & Thoracic Surgery Tags: Original Articles Source Type: research