Spatial stress distribution analysis by thermoelastic stress measurement and evaluation of effect of stress concentration on durability of various orthopedic implant devices

Publication date: 1 June 2017 Source:Materials Science and Engineering: C, Volume 75 Author(s): Yoshimitsu Okazaki, Daisuke Ishii, Atsushi Ogawa Toward the development of highly durable devices, we investigated the effect of the thermoelastic constants of implantable raw metals and the surface stress distribution on the durability of various types of implant device by thermoelastic stress measurement and by evaluating the effect of the stress concentration. Surface stress was dynamically calculated from the bending moment, and the modulus of a section of a device was found to be consistent with the surface stress obtained by thermoelastic stress measurement. The durability limits of various types of bone plate and compression hip screw (CHS) calculated from maximum load vs number of cycles data (L–N data) were close to the notch fatigue strength of the raw material. The concentration factor of an artificial hip stem surface was estimated by comparing the L–N data of the stem and the SN curve of the raw material. The dynamic analysis of durability by thermoelastic stress measurement is useful for selecting the worst case (a product deteriorating to the most severe state) in medical device design.
Source: Materials Science and Engineering: C - Category: Materials Science Source Type: research