Initial clinical experience with ArcCHECK for IMRT/VMAT QA.

Initial clinical experience with ArcCHECK for IMRT/VMAT QA. J Appl Clin Med Phys. 2016;17(5):6118 Authors: Aristophanous M, Suh Y, Chi PC, Whittlesey LJ, LaNeave S, Martel MK Abstract Many devices designed for the purpose of performing patient-specific IMRT/VMAT QA are commercially available. In this work we report our experience and initial clinical results with the ArcCHECK. The ArcCHECK consists of a cylindrical array of diode detectors measuring entry and exit doses. The measured result is a cumulative dose displayed as a 2D matrix. The detector array requires both an absolute dose calibration, and a calibration of the detector response, relative to each other. In addition to the calibrations suggested by the manufacturer, various tests were performed in order to assess its stability and performance prior to clinical introduction. Tests of uniformity, linearity, and repetition rate dependence of the detector response were conducted and described in this work. Following initial test-ing, the ArcCHECK device was introduced in the clinic for routine patient-specific IMRT QA. The clinical results from one year of use were collected and analyzed. The gamma pass rates at the 3%/3 mm criterion were reported for 3,116 cases that included both IMRT and VMAT treatment plans delivered on 18 linear accelera-tors. The gamma pass rates were categorized based on the treatment site, treatment technique, type of MLCs, operator, ArcCHECK device, a...