Transitioning from Triton X-100 to Tergitol 15-S-9: impacts on diagnostic assays using viral PCR  sample solution

Biotechniques. 2024 May 5. doi: 10.1080/07366205.2024.2342172. Online ahead of print.ABSTRACTIn 2019, the European Union banned Triton X-100, a detergent widely used in laboratory diagnostics, including the Viral PCR Sample Solution (VPSS), and urged manufacturers to find environmentally sustainable alternatives. Tergitol 15-S-9 (VPSS2) has been proposed as an alternative surfactant. This multicenter study evaluated the effectiveness of VPSS2, a Tergitol-based viral solution, as a replacement for VPSS. Our results show the equivalent performance of VPSS2 to VPSS for nucleic acid extraction and viral stability over time at different temperatures. The new VPSS formulation was also tested against external quality assurance panels and clinical samples. The results of this work support adopting this modified viral PCR sample solution to replace Triton X-100-containing viral transport solutions.PMID:38706171 | DOI:10.1080/07366205.2024.2342172
Source: BioTechniques - Category: Biotechnology Authors: Source Type: research