In vivo diffusion MRI of the human heart using a 300 mT/m gradient system

CONCLUSION: This work demonstrates cardiac DWI in vivo with higher b-value and higher order of motion compensated diffusion gradient waveforms than is commonly used. Increasing the motion compensation order from M 2 $$ {M}_2 $$ to M 3 $$ {M}_3 $$ and the maximum b-value from 450 to 1000 s / mm 2 $$ \mathrm{s}/{\mathrm{mm}}^2 $$ affected the MD values but FA and the angular metrics (HA and E2A) remained unchanged. Our work paves the way for cardiac DWI on the next-generation MR scanners with high-performance gradient systems.PMID:38650395 | DOI:10.1002/mrm.30118
Source: Magnetic Resonance in Medicine - Category: Radiology Authors: Source Type: research