Validation of intraocular pressure measurement using tonometer AVIA across different postures: A Bland Altman analysis
CONCLUSION: Good reliability exists between IOP measurements using GAT and TPA. However, the devices are not interchangeable and therefore cannot be used reciprocally in the same patient.PMID:38623622 | DOI:10.1177/11206721241247672
Source: European Journal of Ophthalmology - Category: Opthalmology Authors: Sana' Muhsen Alexander Rabadi Mahmoud Alqudah Abdelrahman Obiedat Liyana Owies Ibrahim Alhawaniah Sufian Abdel Hafez Abdallah Al-Ani Source Type: research