Evaluation of a vibration modeling technique for the in-vitro measurement of dental implant stability

This study's purpose was to determine whether changes in density, bonding, and drilling technique affect the measured vibration of a dental implant, and whether they can be quantified as a change in the estimated BII stiffness. Stability was also measured using RFA, insertion torque (IT) and the pullout test. Bone-level tapered implants (4.1 mm diameter, 10 mm length) were inserted in polyurethane foam as an artificial bone substitute. Samples were prepared using different bone densities (20, 30, 40 PCF), drilling sequences, and superglue to simulate a bonded implant. Measurements were compared across groups at a significance level of 0.05. The ASC was able to indicate changes in each factor as a change in the interfacial stiffness. IT and pullout force values also showed comparable increases. Furthermore, the relative difference in ISQ values between experimental groups was considerably smaller than the ASC. While future work should be done using biological bone and in-vivo systems, the results of this in-vitro study suggest that modelling of the implant system with a vibration-based approach may provide a noninvasive method of assessing the mechanical stability of the implant.PMID:38588632 | DOI:10.1016/j.jmbbm.2024.106537
Source: Journal of the Mechanical Behavior of Biomedical Materials - Category: Materials Science Authors: Source Type: research