< em > Technical Note: < /em > Comparison of Alinity c Hemoglobin A1c Immunoassay with Premier Hb9210 Automated HPLC Assay: A Preliminary Report

CONCLUSIONS: We conclude that HbA1c assay on the Alinity c analyzer is a viable alternative to HPLC for measuring HbA1c in clinical laboratories but values at the decision points must be interpreted with caution and if necessary should be repeated by a reference HPLC method.PMID:38514062
Source: Annals of Clinical and Laboratory Science - Category: Laboratory Medicine Authors: Source Type: research