A rapid and sensitive LC-MS/MS method for quantifying oxycodone, noroxycodone, oxymorphone and noroxymorphone in human plasma to support pharmacokinetic drug interaction studies of oxycodone

Biomed Chromatogr. 2024 Apr 8:e5874. doi: 10.1002/bmc.5874. Online ahead of print.ABSTRACTA sensitive and reliable LC-MS/MS method was developed and validated for the quantification of oxycodone and metabolites in human plasma. The method has a runtime of 6 min and a sensitivity of 0.1 μg/L for all analytes. Sample preparation consisted of protein precipitation. Separation was performed on a Kinetix biphenyl column (2.1 × 100 mm, 1.7 μm), using ammonium formate 5 mm in 0.1% aqueous formic acid and methanol LC-MS grade 100% in gradient elution at a flow rate of 0.4 ml/min. Detection was performed in multiple reaction monitoring mode using positive electrospray ionization. The method was linear over the calibration range of 0.1-25.0 μg/L for oxycodone, noroxycodone and noroxymorphone and 0.1-5.0 μg/L for oxymorphone. The method demonstrated good performance in terms of intra- and interday accuracy (86.5-110.3%) and precision (CV 1.7-9.3%). The criteria for the matrix effect were met (CV < 15%) except for noroxymorphone, for which an additional method was applied to compensate for the matrix effect. Whole blood samples were stable for 4 h at room temperature. Plasma samples were stable for 24 h at room temperature and 3 months at -20°C. Furthermore, the method was successfully applied in a pharmacokinetic drug interaction study of oxycodone and enzalutamide in patients with prostate cancer.PMID:38587098 | DOI:10.1002/bmc.5874
Source: Biomedical Chromatography : BMC - Category: Biomedical Science Authors: Source Type: research