Laboratory workflow for optimizing leukocyte count and differential in synovial fluids on Sysmex XN-1000
CONCLUSION: Our workflow based on scattergram interpretation resulted in accurate WBC counts in synovial fluid using automated/and or manual methods. It is important to identify the presence of "banana-shape" scattergrams to avoid overestimated automated WBC counts. Overall, automated synovial WBC count can be used, even for patients with arthroplasty, but after visual inspection of the scattergram to exclude possible interferences.PMID:38530029 | DOI:10.1111/ijlh.14274
Source: International Journal of Laboratory Hematology - Category: Hematology Authors: Marthe Vanrenterghem Julie Dom Nancy Boeckx Melissa Depypere Glynis Frans Georges Vles Christine Van Laer Source Type: research
More News: Hematology | Laboratory Medicine