Study on induced radioactivity and individual dose evaluation in Gantry room for Varian ProBeam Proton Therapy System

This study focuses on systematically investigating the induced radioactivity in the gantry room during pencil beam scanning, utilizing both experimental measurements and Monte Carlo simulations. Results indicate that patients are the primary source of induced radioactivity, predominantly producing radionuclides such as 11C, 13N, and 15O. Long-term irradiation primarily generates radionuclides like 22Na, 24Na, and 54Mn etc. Additionally, this study estimates the individual doses received by medical workers in the gantry room, the irradiation dose for patient escorts, and the additional dose to patients from residual radiation. Finally, the study offers recommendations to minimize unnecessary irradiation doses to medical workers, patient escorts, and patients.PMID:38513476 | DOI:10.1016/j.apradiso.2024.111297
Source: Applied Radiation and Isotopes - Category: Radiology Authors: Source Type: research