Anal Chim Acta. 2024 Apr 15;1298:342401. doi: 10.1016/j.aca.2024.342401. Epub 2024 Feb 26.ABSTRACTBACKGROUND: High-resolution matrix-assisted laser desorption/ionization-time of flight mass spectrometry (MALDI-TOF MS) and nuclear magnetic resonance (NMR) spectroscopy are powerful tools to identify unknown psychoactive substances. However, in complex matrices, trace levels of unknown substances usually require additional fractionation and concentration. Specialized liquid chromatography systems are necessary for both techniques. The small flow rate of nano LC, typically paired with MALDI-TOF MS, often results in prolonged fractionation times. Conversely, the larger flow rate of semi-preparative LC, used for NMR analysis, can be time-consuming and labor-intensive when concentrating samples. To address these issues, we developed an integrated automatic system that integrated to regular LC.RESULT: Automatic spot collector (ASC) and automatic fraction collector (AFC) were present in this study. The ASC utilized in-line matrix mixing, full-contact spotting and real time heating (50 °C), achieving great capacity of 5 μL droplet on MALDI plate, high recovery (76-116%) and rapid evaporation in 2 min. The analytes were concentrated 4-8 times, forming even crystallization, reaching the detection limit at the concentration of 50 μg L-1 for 12 psychoactive substances in urine. The AFC utilizes flexible tubing which flash-tapped the microtube's upper rim (3 mm depth) instead of reaching...