COVID-19 electrochemical immunosensor with Ag-MOF: Rapid and high-selectivity nasal swab testing for effective detection

In this study, we leveraged the properties of synthesized Ag-MOF, including high porosity and increased flow intensity. Electrochemical techniques such as cyclic voltammetry (CV) and differential pulse voltammetry (DPV) were employed to develop an economical and portable sensor with exceptional selectivity for COVID-19 detection. The methodology involves the deposition of Ag-MOF onto the surface of a Glassy Carbon Electrode (GCE), which resulted in a progressive augmentation of electric current. Subsequently, the targeted antibodies were applied, and relevant tests were conducted. The sensor demonstrated the capacity to detect the virus within a linear range of 100 fM to 10 nM, boasting a noteworthy Limit of Detection (LOD) of 60 fM. The entire detection process could be completed in a brief duration of 20 min, exhibiting high levels of accuracy and precision, outperforming comparable techniques in terms of speed and efficacy.PMID:38431139 | DOI:10.1016/j.ab.2024.115500
Source: Analytical Biochemistry - Category: Biochemistry Authors: Source Type: research