Evaluation of a novel patient-specific quality assurance phantom for robotic single-isocentre, multiple-target stereotactic radiosurgery, and stereotactic radiotherapy
CONCLUSIONS: A novel spherical phantom was evaluated for the CyberKnife system and achieved acceptable PSQA passing rates using TG218 recommendations. The phantom can measure true-composite dose and offers high-resolution results for PSQA, making it a valuable device for robotic radiosurgery.ADVANCES IN KNOWLEDGE: This is the first study on PSQA of 3 targets concurrently on the CyberKnife system.PMID:38401536 | DOI:10.1093/bjr/tqae011
Source: The British Journal of Radiology - Category: Radiology Authors: Qianyi Xu Yevgeniy Vinogradskiy Jimm Grimm Wei Nie Pamela Dupre Ashish K Chawla Gopal Bajaj Haihua Yang Tamara LaCouture Jiajin Fan Source Type: research