This study aimed to improve the performance of single-channel steady-state visual evoked potential (SSVEP)-based visual acuity assessment by mode decomposition methods. Using the SSVEP dataset induced by the vertical sinusoidal gratings at six spatial frequency steps from 11 subjects, 3-40-Hz band-pass filtering and other four mode decomposition methods, i.e., empirical mode decomposition (EMD), ensemble empirical mode decomposition (EEMD), improved complete ensemble empirical mode decomposition with adaptive noise (ICEEMDAN), and variational mode decomposition (VMD), were used to preprocess the single-channel SSVEP signals from Oz electrode. After comparing the SSVEP signal characteristics corresponding to each mode decomposition method, the visual acuity threshold estimation criterion was used to obtain the final visual acuity results. The agreement between subjective Freiburg Visual Acuity and Contrast Test (FrACT) and SSVEP visual acuity for band-pass filtering (−0.095 logMAR), EMD (−0.112 logMAR), EEMD (−0.098 logMAR), ICEEMDAN (−0.093 logMAR), and VMD (−0.090 logMAR) was all pretty good, with an acceptable difference between FrACT and SSVEP acuity for band-pass filtering (0.129 logMAR), EMD (0.083 logMAR), EEMD (0.120 logMAR), ICEEMDAN (0.103 logMAR), and VMD (0.108 logMAR), finding that the visual acuity obtained by these four mode decompositions had a lower limit of agreement and a lower or close difference compared to the traditional band-pass filtering met...