Computer-based analysis of the taper connection strength of different revision head and adapter sleeve designs

CONCLUSION: A sufficient assembly force should be ensured intraoperatively, especially with an increasing head diameter and adapter sleeve size, as lower assembly forces might lead to reduced taper connection strength.PMID:37698840 | DOI:10.1515/bmt-2023-0066
Source: Biomedizinische Technik/Biomedical Engineering - Category: Biomedical Engineering Authors: Source Type: research