Interlaboratory evaluation of quality control methods for circulating cell-free DNA extraction

In this study we evaluated the application of nucleic acid "spike-in" control materials to aid quality control (QC) and standardisation of cfDNA isolation for use in in vitro diagnostic assays. We describe an approach for the design and characterisation of in-process QC materials, illustrating it with a spike-in material containing an exogenous Arabidopsis sequence and DNA fragments approximating to ccfDNA and genomic DNA lengths. Protocols for inclusion of the spike-in material in plasma ccfDNA extraction and quantification of its recovery by digital PCR (dPCR) were assessed for their suitability for process QC in an inter-laboratory study between five expert laboratories, using a range of blood collection devices and ccfDNA extraction methods. The results successfully demonstrated that spiking plasmid-derived material into plasma did not deleteriously interfere with endogenous ccfDNA recovery. The approach performed consistently across a range of commonly-used extraction protocols and was able to highlight differences in efficiency and variability between the methods, with the dPCR quantification assay performing with good repeatability (generally CV <5%). We conclude that initial findings demonstrate that this approach appears "fit for purpose" and spike-in recovery can be combined with other extraction QC metrics for monitoring the performance of a process over time, or in the context of external quality assessment. AVAILABILITY OF DATA AND MATERIALS: Processed data is...
Source: New Biotechnology - Category: Biotechnology Authors: Source Type: research