Video plethysmography for contactless measurement of respiratory rate in surgical patients

The objective of our study was to determine whether VPPG can feasibly and accurately measure RR without contact in surgical patients at a clinical setting.  After research ethics approval, 216 patients undergoing ambulatory surgery consented to the study. Patients had a 1.5 min video of their faces taken via an iPad preoperatively, which was analyzed using VPPG to obtain RR information. The RR prediction by VPPG was compared to 60-s manual counting o f breathing by research assistants. We found that VPPG predicted RR with 88.8% accuracy and a bias of 1.40 ± 1.96 breaths per minute. A significant and high correlation (0.87) was observed between VPPG-predicted and manually recorded RR. These results did not change with the ethnicity of patie nts. The success rate of the VPPG technology was 99.1%. Contactless RR monitoring of surgical patients at a hospital setting using VPPG is accurate and feasible, making this technology an attractive alternative to the current approaches to RR monitoring. Future developments should focus on improvin g reliability of the technology.
Source: Journal of Clinical Monitoring and Computing - Category: Information Technology Source Type: research