Assessment of resolution and noise in magnetic resonance images reconstructed by data driven approaches

CONCLUSION: Objective image quality parameters, local resolution and g-factors have been determined. The examined data driven methods show less local blurring than compressed sensing. The noise enhancement for reconstructions using CS, VN and U-Net is elevated at anatomical contours but is drastically reduced with respect to GRAPPA. Overall, the applied framework provides the possibility for more detailed analysis of novel reconstruction approaches incorporating non-linear and non-stationary transformations.PMID:37684119 | DOI:10.1016/j.zemedi.2023.08.007
Source: Zeitschrift fur Medizinische Physik - Category: Radiology Authors: Source Type: research