Glucose measurements with accu check inform II versus hexokinase plasma method during surgery under general anesthesia, an observational cohort study

ConclusionThe agreement between glucose measurements using ACI II and the reference laboratory hexokinase test was clinically acceptable with a percentage error of 10.0% (95% CI 8.0 to 11.9). The use of TIVA may negatively affect the measurement performance of the ACI II.
Source: Journal of Clinical Monitoring and Computing - Category: Information Technology Source Type: research