Long-term experience in quality assurance of on-rail computed tomography systems for image-guided radiotherapy using in-house multifunctional phantoms

This study reported the QA status of the system during the period 2014–2021. The absolute mean accuracy of couch rotation was 0.4 ± 0.28 mm, 0.44 ± 0.36 mm, and 0.37 ± 0.27 mm in th e SI, RL, and AP directions, respectively. Horizontal and remote movement accuracies of the treatment couch were also within 0.5 mm of the absolute mean value. A decrease in the accuracy of couch rotation was also observed due to aging deterioration of related parts caused by the frequent use of co uch rotation. The three-dimensional accuracy of on-rail CT systems derived mainly from treatment couches can be maintained within 0.5 mm with appropriate accuracy assurance for at least >  8 years.
Source: Radiological Physics and Technology - Category: Physics Source Type: research
More News: CT Scan | Physics | Study