Validation of an ex vivo Flow Model Including Magnetic Resonance Imaging to Study the Effects of Endovascular Treatments on the Arterial Wall

This study developed and validated an ex vivo flow model using porcine iliac arteries, obtained from a local abattoir. Twenty arteries (of 10 pigs) were equally allocated to two groups: a mock-treated control group and an endovascular intervention group. Arteries of both groups were perfused with porcine blood for 9 min, including 3 min of balloon angioplasty in the intervention group. Vessel injury was assessed by calculating the presence of endothelial cell denudation, vasomotor function, and histopathological analysis. MR imaging displayed balloon positioning and inflation. Endothelial cell staining showed 76% of denudation after ballooning compared to 6% in the control group (p #x3c; 0.001). This was confirmed by histopathological analysis, showing a significantly reduced endothelial nuclei count after ballooning compared to the controls (median: 22 vs. 37 nuclei/mm,p = 0.022). In the intervention group, vasoconstriction and endothelium-dependent relaxation were significantly reduced (p #x3c; 0.05).We present an ex vivo flow model to test the effects of endovascular therapy on the vessel ’s wall morphology, endothelial denudation, and endothelial-dependent vasomotor function under physiological conditions. Additionally, it allows the future testing of human arterial tissue.J Vasc Res 2023;60:1 –11
Source: Journal of Vascular Research - Category: Research Source Type: research