Monte Carlo simulation study to explore optimum conditions for Astatine-211 SPECT

Abstract211At is a promising nuclide for targeted radioisotope therapy. Direct imaging of this nuclide is important for in vivo evaluation of its distribution. We investigated suitable conditions for single-photon emission computed tomography (SPECT) imaging of211At and assessed their feasibility using a homemade Monte Carlo simulation code, MCEP-SPECT. Radioactivity concentrations of 5, 10, or 20  kBq/mL were distributed in six spheres in a National Electrical Manufactures Association (NEMA) body phantom with a background of 1 kBq/mL. The energy window, projection number, and acquisition time were 71–88 keV, 60, and 60 s, respectively, per projection. A medium-energy collimator and thr ee low-energy collimators were tested. SPECT images were reconstructed using the ordered subset expectation maximization (OSEM) method with attenuation correction (Chang method) and scatter correction (triple-energy-windows method). Image quality was evaluated using the contrast-to-noise ratio (CNR) for detectability and the contrast recovery coefficient (CRC) for quantitavity. The low-energy, high-sensitivity collimator exhibited the best detectability among the four types of collimators, with a maximum CNR value of 43. In contrast, the low-energy, high-resolution collimator exhibited excelle nt quantitavity, with a maximum CRC value of 102%. Scatter correction improved the image quality. In particular, the CRC value almost doubled after scatter correction. The detection of spheres smal...
Source: Radiological Physics and Technology - Category: Physics Source Type: research