Development of a device that remotely removes a mask in the head and neck immobilization system: a prototype and demonstration experiment

In this study, a prototype device was developed to quickly remove the mask used to immobilize the head and neck by remotely releasing the quick fasteners. As a first step in investigating the usefulness of this prototype, we performed repeated removal tests and examined the accuracy of dose calculation. The results showed that the quick-release fasteners of a Type-S system (CIVCO Medical Solutions, Iowa, USA) could be removed remotely and accurately (success rate: 100%). Additionally, the dose errors in treatment planning were negligible (<  1.0%), and the gamma pass rate was equivalent (99.9%). Therefore, this prototype device with a remote system would help manage patient safety in emergencies, such as a disaster or a sudden change in the patient's condition. However, age-related deterioration with long-term clinical use or its abi lity to link with beam-off still requires further exploration.
Source: Radiological Physics and Technology - Category: Physics Source Type: research