Modified AngioVac System use With Extracorporeal Membrane Oxygenation in a Child With Fontan Thrombosis

We report modified use of the second-generation AVS with concurrent venoarterial extracorporeal membrane oxygenation (VA-ECMO) in a 10-year-old, 23 kg patient with failing Fontan circulation due to acute-on-chronic Fontan and PA thrombosis. Emergent femoral VA-ECMO was initiated for profound hypoxemia during cardiac catheterization. Fluoroscopy-guided thrombo-embolectomy was performed with AVS components. A two-venous limb inflow cannulation strategy was used; the AngioVac cannula (AVC) was positioned in the central PA and joined to the existing ECMO venous-limb. Thrombus extraction proceeded by clamping the ECMO venous-limb with the existing VA-ECMO centrifugal pump generating negative pressure to the AVC. Extracorporeal membrane oxygenation arterial-limb access was used for blood return. Settings to avoid cavitation were a maximal negative pressure of 100 mmHg using
Source: ASAIO Journal - Category: Medical Devices Tags: Case Reports Source Type: research