Sensors, Vol. 21, Pages 5056: Passive Electrical Damping of a Quartz Tuning Fork as a Path to Fast Resonance Tracking in QEPAS

Sensors, Vol. 21, Pages 5056: Passive Electrical Damping of a Quartz Tuning Fork as a Path to Fast Resonance Tracking in QEPAS Sensors doi: 10.3390/s21155056 Authors: Roman Rousseau Diba Ayache Wioletta Trzpil Michael Bahriz Aurore Vicet In Quartz-Enhanced PhotoAcoustic Spectroscopy (QEPAS) gas sensors, the acoustic wave is detected by the piezoelectric Quartz Tuning Fork (QTF). Due to its high-quality factor, the QTF can detect very low-pressure variations, but its resonance can also be affected by the environmental variations (temperature, humidity, …), which causes an unwanted signal drift. Recently, we presented the RT-QEPAS technique that consistently corrects the signal drift by continuously measuring the QTF resonance. In this article, we present an improvement of RT-QEPAS to fasten the QTF characterization time by adding a passive electronic circuit, which causes the damping of the QTF resonance. The damping circuit is optimized analytically and through SPICE simulation. The results are supported by experimental observations, showing a 70 times improvement of the relaxation times compared to the lone QTF, which opens the way to a fast and drift-free QEPAS sensor.
Source: Sensors - Category: Biotechnology Authors: Tags: Article Source Type: research