Reducing contrast dose using virtual monoenergetic imaging for aortic CTA.

Reducing contrast dose using virtual monoenergetic imaging for aortic CTA. J Appl Clin Med Phys. 2020 Jul 02;: Authors: Yoshida R, Usui K, Katsunuma Y, Honda H, Hatakeyama K Abstract Three-dimensional computed tomographic angiography (3D-CTA) is widely used to evaluate the inner diameters of vessels and the anatomical vascular structure prior to endoscopic aortic surgery or transcatheter valve implantation. Virtual monoenergetic imaging (VMI) is a new application in dual-energy CT (DECT). We evaluated the potential for contrast dose reduction in preoperative aortic CTA using VMI. To evaluate performance in terms of image quality and vessel shape, we quantified the contrast-to-noise ratio (CNR) and the vessel diameter using a cylinder phantom we developed, and used volume rendering to assess visual quality. All VMI had improved CNR values compared with conventional 120 kVp images at an iodine content of 15 mgI/mL. In each image, a virtual mono-energy of 40 keV yielded the highest CNR value, and an iodine content of 9 mgI/mL was comparable to that of conventional images with an iodine content of 15 mgI/mL. The circularity indices (CI) of the vascular model at 15, 12, and 9 mgI/mL were similar to those of the reference condition using conventional voltages; however, CI was degraded at iodine contents of 6 and 3 mgI/mL with VMI. In the case of iodine content of 15 mgI/mL, VMI was superior, with conventional image by visual evalua...
Source: Journal of Applied Clinical Medical Physics - Category: Physics Authors: Tags: J Appl Clin Med Phys Source Type: research