Early Mechanical Failure of a Tumoral Endoprosthesic Rotating Hinge in the Knee: Does Bumper Wear Contribute to Hyperextension Failure?

CONCLUSIONS: The Zimmer Segmental rotating-hinge tumoral prosthesis underwent revision for implant breakage at short term in three of six patients after tumor resection and reconstruction of the knee. Bumper breakage was associated with posterior instability that was related to wear of the bushing blocking system. We are unaware of reports of these issues by other observers or in other prosthetic designs, but we feel larger registries should be created to see if this failure mechanism has been observed by others. If so, this design needs to be improved or the blocking system should be avoided.Level of Evidence Level IV, therapeutic study. PMID: 31764341 [PubMed - in process]
Source: Clinical Orthopaedics and Related Research - Category: Orthopaedics Authors: Tags: Clin Orthop Relat Res Source Type: research