Triboelectric charging of polytetrafluoroethylene antithrombotic catheters

This study proposes that a polytetrafluoroethylene (PTFE) electret tube charged by frictional electricity can prevent the solidification of the indwelling catheter in blood vessels. Coagulation in intravascular indwelling catheters may discontinue the treatment because of thrombus-derived bacteria –adhesion infections or poor blood removal. Current commercially available intravascular catheters lack complete antithrombotic measures, even with heparin or urokinase antithrombotic coatings. Herein, we tested the effectiveness of an antithrombotic treatment that prevents coagulation using a sta tic electric charge on the interior of the PTFE tube via the triboelectric effect by rubbing the tube’s inner wall with a round glass rod. The anticoagulation properties were evaluated by enclosing a sample of blood in an electret tube and observing the coagulase adhering to the inner wall using a microscope. To confirm the effectiveness of this treatment, the charge–distribution on the inner surface of the electret tube was measured, surface irregularities were observed, and the elements on the surface were analyzed. The surface potential inside the electret tube was − 366.4 V, which proved effective for an antithrombotic treatment, as it discouraged coagulation, and the triboelectric charging process caused neither surface element denaturation nor significant surface irregularities. The nearly uniform negative surface charge on the inside of the tube was responsible for the an...
Source: Journal of Artificial Organs - Category: Transplant Surgery Source Type: research