Minimally Invasive Proximal Interphalangeal Joint Arthrodesis Using Headless Screw: Surgical Technique

We present a minimally invasive PIP arthrodesis that provides rigid fixation with a headless compression screw. Seven patients who presented to the senior author with PIP joint arthritis underwent PIP arthrodesis by minimally invasive technique. A 1 cm transverse incision is made over the PIP joint, incising skin, central band, and articular capsule. PIP joint is flexed to expose the articular surface. Articular surfaces are prepared with a fine tipped rongeur, exposing subchondral bone until flat surfaces are obtained. Under fluoroscopy a guide wire for cannulated headless screw (3.0, 2.4, or 2.0 mm) is inserted in an antegrade manner. It progresses from the center of the proximal phalangeal articular surface until it exits through the dorsal cortex and the distal end lies within the subchondral bone. This is the most critical step of the procedure because the guide wire angle determines the degree of flexion of the fusion. A 5 mmincision is made over the guide wire and the wire is advanced through the center of the medullary canal of the middle phalanx. The wire is then overdrilled, length is measured, and a headless compression screws is inserted. Reevaluate alignment after insertion of the screws because malrotation may be induced by torque during compression. Six consecutive patients underwent the procedure by the senior author. All patients healed the arthrodesis without complications and hardware removal was not needed. Minimally invasive PIP joint arthrodesis is...
Source: Techniques in Hand and Upper Extremity Surgery - Category: Surgery Tags: Techniques Source Type: research