Asynchrony Between Ventilator Flow and Pressure Waveforms and the Capnograph on Dräger Anesthesia Workstations: A Case Report

Modern anesthesia workstations display capnography, flow-time, and pressure-time waveforms in real time. We observed that at certain ventilator settings (10 breaths/min) on Dräger workstations, the expiratory phase of the capnograph overlaps both the inspiratory and the expiratory phases of ventilation. This discrepancy disappears at respiratory rates of 16 breaths/min. This synchronous respiratory monitoring display at respiratory rates 16 breaths/min is not physiologically correct, because it implies a synchronization of waveforms that is not actually present. This again becomes asynchronous once the respiratory rate is increased to>18 breaths/min. Such an artifact may not affect the patient’s safety in most cases but may mislead clinicians when synchrony between flow/pressure and capnography is needed for diagnostic purposes. We wish to share this discrepancy with clinicians and notify the manufacturer so that potential solutions may be found.