Comparison of aspiration catheter performance using adaptive pulsatile aspiration in an < em > in vitro < /em > thrombectomy model
CONCLUSION: APA via the Von Vascular, Inc. ALGO® pump achieved a high success rate in an in vitro MT model. All catheters tested with the pump achieved complete reperfusion in all trials, and complete clot ingestion into the pump was seen in a majority of trials. The promising in vitro performance of APA using multiple catheters warrants future in vivo investigation.PMID:38693768 | DOI:10.1177/15910199241250082
Source: Interventional Neuroradiology - Category: Radiology Authors: Michael A Silva Sai Sanikommu Pedro Bartkevitch Rodrigues Manning Hanser Ryan Ortiz Victor Gamez Robert M Starke Source Type: research