Paper Spray Mass Spectrometry with On-Paper Electrokinetic Manipulations: Part- Per-Trillion Detection of Per/Polyfluoroalkyl Substances in Water and Opioids in Urine

Angew Chem Int Ed Engl. 2024 Apr 24:e202401729. doi: 10.1002/anie.202401729. Online ahead of print.ABSTRACTWe developed a simple, paper-based device that enables sensitive detection by mass spectrometry (MS) without solid phase extraction or other sample preparation. Using glass fiber filter papers within a 3D printed holder, the device employs electrokinetic manipulations to stack, separate, and desalt charged molecules on paper prior to spray into the MS. Due to counter-balanced electroosmotic flow and electrophoresis, charged analytes stack on the paper and desalting occurs in minutes. One end of the paper strip was cut into a sharp point and positioned near the inlet of a MS. The stacked analyte bands move toward the paper tip with the EOF where they are ionized by paper spray. The device was applied to analysis of PFAS in tap water with sub part-per-trillion detection limits in less than ten minutes with no sample pretreatment. Analysis of opioids in urine also occurs in minutes. The crucial parameters to enable stacking, separation, and MS ionization of both positively and negatively charged analytes were determined and optimized. Experimental and computational modeling studies confirm the electrokinetic stacking and analyte transport mechanisms. On-paper separations were carried out by stacking analyte bands at different locations depending on their electrophoretic mobility, achieving baseline separation in some cases.PMID:38657037 | DOI:10.1002/anie.202401729
Source: Angewandte Chemie - Category: Chemistry Authors: Source Type: research