High-sensitivity troponin testing at the point of care for the diagnosis of myocardial infarction: a prospective emergency department clinical evaluation

In the ED, laboratory high-sensitivity cardiac troponin (hs-cTn) tests facilitate ‘rule-out’ of myocardial infarction (MI).1 Contemporary point-of-care (POC) troponin tests have inferior analytical sensitivity,2 preventing rapid ‘rule-out’ strategies in near-patient settings. The new Siemens Atellica VTLi POC test meets hs-cTn criteria.3 Between June and September 2022, adult ED patients (≥16 years) with symptoms suspicious for MI and no ST-segment elevation on ECG were eligible for inclusion if staff trained on the POC instrument were available. Simultaneously, blood was taken for near-patient testing with the VTLi POC hs-cTnI assay (limit of detection (LOD) of 1.2 ng/L, sex-specific 99th percentile upper reference limits (URLs) of 18 ng/L (female) and 27 ng/L (male), and an optimised ‘rule-out’ threshold of <4 ng/L)3 and for laboratory testing with the Abbott Alinity hs-cTnI assay (LOD of 1.6 ng/L, URLs of 16 ng/L (female) and 34 ng/L (male), and an optimised ‘rule-out’ threshold of <5 ng/L),
Source: Emergency Medicine Journal - Category: Emergency Medicine Authors: Tags: Research letter Source Type: research